GD6900 Capacitance & Tan Delta Tester

GD6900 Capacitance & Tan Delta Tester

Brief Description:

GD6900 measures the capacitance and dielectric loss factor (tgδ) of high voltage electric equipment. It is integrated structure, built-in dielectric loss test bridge, variable frequency adjustable power supply, boosting transformer and SF6 standard capacitor.


Product Detail

Product Tags

Features

• 7-inch touch LCD screen display.

•  Calendar chip and large storage inside. Save testing result according to time order, check history record and print the result.

•  The instrument data can be exported through U disk.

•  The integrated insulation resistance test module can test polarization index, absorption ratio and insulation resistance.

• Multiple testing mode, with modes of internal high voltage, external high voltage, internal standard, external standard, GST/UST, self-excitation. High voltage(more than 10kV) dielectric loss test can be made in the situation of external standard outside high voltage.

•  Test full sealed CVT (Capacitive Voltage Transformer) C1 and C2 dielectric loss and capacitance at the same time. Also test CVT transformation ratio and voltage angle difference.

• No need to dismantle cable to measure dielectric loss and capacitance of CVT.

•  The dielectric loss and capacitance value of C0 in the upper end of CVT can be measured by using the reverse shielding method.

• High speed sampling signal. Inverter and sampling circuit inside are digitized controlled. Output voltage is adjusted continuously.

• LCR automatic measurement. Inductance, capacitance, resistance can be measured and displayed.

• Multiple protection of input voltage fluctuation, output short circuit, over-voltage, over-current, temperature, secure and reliable. Meanwhile, it has the function of grounding testing, that voltage boost is not permitted for non-grounding equipment.

•  Frequency can be changed into 45Hz, 50Hz, 55Hz, 47.5Hz/52.5Hz, 45Hz/55Hz, 60Hz, 65Hz, 57.5Hz/62.5Hz, 55Hz/65Hz.

Specifications

Working condition

-15℃~40℃

RH<80%

Anti-interference principle

Frequency conversion

Power supply

AC 220V, 60Hz

Generator can be used.

High voltage output

0.5kV~10kV

Every 0.1kV

Accuracy

2%

Max. current

200mA

Capacity

2000VA

Self-excitation power

AC 0V~50V/15A

Single frequency: 45.0Hz, 50.0Hz

55.0Hz, 60.0Hz, 65.0Hz

Automatic dual frequency conversion

45.0Hz/55.0Hz, 47.5Hz/52.5Hz

55.0Hz/65.0Hz, 57.5Hz/62.5Hz

Resolution

tgδ: 0.001%

Cx: 0.001pF

Accuracy

∆tgδ: ±(reading*1.0%+0.040%)

∆C x: ±(reading*1.0%+1.00PF)

Measurement range

tgδ

Without limit

Cx

 

15pF < Cx < 300nF

10kV

Cx < 60nF

5kV

Cx < 150nF

1kV

Cx < 300nF

CVT test

Cx < 300nF

LCR measurement range

L>20H(2kV)

R>10KΩ(2kV)

LCR measurement accuracy

2%

Angle resolution

0.01

CVT ratio range

10~10000

CVT ratio accuracy

3%

CVT ratio resolution

0.01

Insulation resistance

DC: 2.0-10kV   accuracy: ±(reading×2%+0.1kV)

10MΩ-10GΩ: ±(reading*5%+0.5M)

10GΩ-100GΩ: ±(reading*5%+0.2G)

100GΩ-1TGΩ: ±(reading*10%+0.5G)

Dimension

Main unit: 580(L)×380(W)×380(H)

Accessory box: 430(L)×320(W)×140(H)

Memory capacity

100 groups, USB flash disk storage is supported.

Weight

Main unit: 23.5Kg

Accessory box: 5.25Kg